Move the cursor down to 'Setup Events' and press the right arrow key or [OK]. The Setup
Events screen will be displayed.
Setup Events
Rate:000:01:00.0
Level +/-:00.100M
Temp +/-:05.0
°
C
Using the cursor keys, set the rate (time interval) at which the event should be monitored.
This value must be more regular than the normal logging rate. In the example above, the
level and temperature will be monitored every minute.
Next set the trigger point for the event logging, in other words, the amount each value must
change from the previously logged value in order to be logged.
So, in the example above, the level and temperature will be monitored every minute, but
will only be logged if the level varies by more than 10cm from the previous logged value or
the temperature varies by more than 5.0°C from the previously logged value.
When using a -CTD version, the Temperature event is replaced by a Salinity event. This is
extremely useful when conducting salt water intrusion studies in coastal regions.
Finally, as with all settings, once the event logging criteria have been set up, transfer the
settings to the Logger by pressing the OK key.
After successful setup of the event logging, the 'ACTION SUCCESSFUL' message will be
displayed. Use the [ESC] key to leave this page after transfer. If you press the [ESC] key
before pressing [OK], your revised settings will not be transferred to the Logger.
4.11.4. Deployment Site Ident Setup
To setup the Probe's Site Ident (a unique 16 character site identifier), press the
MENU
key
to get to the Main Menu, then choose
Probe Deployment
followed by
Set ID & Position.
The Set ID & Position screen will be displayed.
Set ID & Position
[Use arrow keys]
ID:Borehole 1
View/Sync Position
The sixteen characters to the right of the 'ID:' label are the current site ident loaded into the
connected Logger.
To alter the Site Ident, input the new site name (up to sixteen characters) using the
up/down and left/right arrow keys.
As with all settings, once the new site ID has been setup, transfer the data to the Logger by
pressing the OK key. If you press the [ESC] key from this screen, your revised settings will
not be transferred to the Logger.