6. Connecting to Third Party Telemetry and Logging Devices
6.1. Introduction
The LeveLine
®
range of probes are designed to output live pressure and temperature data
to a third party logging device using industry standard interfaces. The -CTD version also
outputs live electrical conductivity and salinity data.
In order to translate the live pressure data into water level above the sensor, it is necessary
to perform further calculations. These calculations are automatically carried out in the
Aquaread
®
LeveLine
®
Meter and LeveLink
™
PC Software, but if the LeveLine
®
probe is
connected to a third party logging or telemetry device, level (depth) must be calculated by
that device.
6.17. Depth Calculation for Third Party Devices
for further details.
6.2. Connections
The LeveLine
®
accessory cable LL-EXT-TEL-AD (extension cable to flying lead adaptor for
telemetry) allows any LeveLine
®
Probe to be connected, via a standard LeveLine
®
extension cable (EXT-LL-XM), to a third party controller device.
This accessory terminates with a screened multicore cable, which allows external power to
be supplied to the LeveLine
®
and facilitates bi-directional serial data communication. The
basic model LeveLine
®
Mini Probe has a multicore cable permanently attached.
When installing a gauge (vented) type LeveLine
®
probe with a vented cable, it is essential to
fit a desiccant housing and cartridge to the open end of the cable in order to prevent any
moisture entering the cable or probe. See section
3.2.6.Fitting a Desiccant Housing and