Table 3 summarises the Modbus functions supported by the LeveLine
®
and the following
subsections describe their command & response packet formats.
Function
Function Code
Read Holding Registers
0x03
Read Input Registers
0x04
Write Holding Register
0x06
Write Holding Registers
0x10
Report Slave ID
0x11
Table 3: Supported Modbus Functions
6.14.1. Exception Responses
If a command is unsupported or has invalid parameters, the LeveLine
®
will return an
exception response with the format:
Field
Size
Content
<a>
1 byte
Slave address
<exception-function-
code>
1 byte
Function code of the c 0x80
<exception-code>
1 byte
The type of exception
1 – Unsupported command function code
2 – Invalid data address in command
3 – Invalid data value in command
<crc>
2 bytes
CRC
6.14.2. Read Holding Registers
This command reads one or more Modbus holding registers, and has the format:
Field
Size
Content
<a>
1 byte
Slave address
<function-code>
1 byte
0x03
<start-address>
2 bytes
Base address of the block of registers to be read
(0x0000-0xFFFF)
<nr-registers>
2 bytes
The number of registers to be read (0x0001-
0x007D)
<crc>
2 bytes
CRC
NOTE: The address in the <start-address> field is the register number – 1.
The LeveLine
®
will send a response with the format:
Field
Size
Content
<a>
1 byte
Slave address
<function-code> 1 byte
0x03
<values-size>
1 byte
Size in bytes of <values> field (equal to 2 x <nr-
registers>)
<values>
<values-
size> bytes
Block of data containing the values of the requested
registers, 2 bytes per register, registers in order of
increasing address from <start-address> onwards
<crc>
2 bytes
CRC