LeveLine
®
Instruction Manual
10401-01026 Rev D
Apply a generous smear of grease to the connector O-rings situated just inside the
moveable retaining collar on the plug. Be careful not to get any grease inside the connector
near the gold contacts. A small smear of grease should also be applied to the thread on the
Logger to allow easy tightening of the collar.
To connect the Extension Cable to the LeveLine
®
Logger, align the spot on the LeveLine
®
Logger with the
◄AQUAREAD
logo on the plug body, then press the plug into the Logger
and tighten the retaining collar fully.
DO NOT TWIST THE CONNECTOR BODY WITH
RESPECT TO THE LOGGER
Once a Logger is connected to the LeveLine
®
Meter, switch the Meter on by briefly pressing
the red on/off switch. The LeveLine
®
Meter should detect the Logger and automatically
display the Logger’s serial number, firmware revision and pressure rating.
4.6. Taking Measurements
After a few seconds, the Meter will automatically start to display live readings along with the
current GPS status. The initial data screen for the LeveLine
®
Meter in conjunction with a
Level Logging Probe is shown below. If connected to a Baro Logger, lines three and four
will be blank.
BARO:1013mb
3
D
TEMP:012.53
°
C
LEVEL:006.048M
Hit [OK] to zero
When connected to a Level Logging Probe, the BARO value on the top line is derived from
the Meter's internal barometric sensor. When connected to a Baro Logger, the BARO value
is derived from the barometric sensor in the Logger.
When connected to a Level measuring Probe, the [OK] key can be used to zero the level
reading, as indicated on the bottom line of the display. Pressing [OK] will apply an offset to
the current level reading to make it read zero on the Meter. All subsequent level readings
on the Meter will be subject to this offset until the Meter's power is cycled or the level is
zeroed again.
This means that subsequent levels may be either positive or negative with respect to the
depth at which the reading was zeroed. Zeroing the level in this way does not have any
affect on the Logger or the logged data. It is a function of the Meter only.
Left/right arrows at the bottom corners of the screen indicate further data screens are
available. To access these screens, simply press either the left or right arrow keys. Any
value that is out of range or unavailable will be displayed as dashes. The other screens
available are shown below.
PROBE STATUS
3
D
Records:006216
Battery:96.0%
Logging:NO
Probe Status Screen
When connected to a Mini version, the values displayed on this screen will not be valid as
there is no memory, battery or logging facility.