LeveLine
®
Instruction Manual
10401-01026 Rev D
rest of the packet (currently 0x01)
<data>
12 bytes (format
version 1)
Block of data containing product information
<crc>
2 bytes
CRC
Format version 1 of the <data> block is:
<serial-nr>
8 bytes
The LeveLine's serial number (an ASCII string)
<fw-ver>
2 bytes
The LeveLine's firmware version number x 100
<pressure-sensor-
type>
1 bytes
The pressure sensor type (0 = absolute, 1 =
gauge)
<pressure-rating>
1 bytes
The pressure sensor rating (bar)
6.15. MODBUS Register Map
The Modbus specification defines 4 classes of addressable data: discrete inputs, coils,
input registers & holding registers. The characteristics of these classes are shown in the
table below.
Read Only
Read/Write
Single
bit
Discrete
Inputs
Coils
16-bit
Input
Registers
Holding
Registers
Modbus Data Classes
The LeveLine
®
implements all of its registers as holding registers or input registers. The
single bit classes are not used and the command functions for accessing them are not
implemented.
The holding registers allowing configuration of the LeveLine
®
are summarised in the table
below.
Register
Number
Register Address (used in
Modbus commands)
Description
1
0x0000
Slave address (1 – 247)
2
0x0001
Modbus mode (0 = RTU, 1 =
ASCII)
3
0x0002
Serial communication speed
(1200, 2400, 4800, 9600, 19200,
38400, 57600)
4
0x0003
Serial communication parity (0 =
none, 2 = even, 3 = odd)
Modbus Holding Registers
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 81 of 90