LeveLine
®
Instruction Manual
10401-01026 Rev D
1.4.
LeveLine-CTD
®
Battery Life
All LeveLine-CTD
®
Loggers run from a single, dealer replaceable 3.6V Lithium-thionyl
Chloride (Li-SOCL
2
) battery.
The table below shows typical battery life of the LeveLine-CTD
®
Loggers at various
logging/event checking rates at 25°C. Higher temperatures can increase the rate of self
discharge and hence reduce the figures shown.
Logging / Event Checking Interval
Expected Battery Life
1 Sec
6 Days
15 Sec
90 Days
1 Min
1 Year
5 Min
5 Years
10 Min
10 Years
> 10 Min
> 10 Years
1.4.1. Important Note Regarding Battery Life
If you set a long logging interval but a very short event checking interval, the battery life
should be estimated on the
event checking interval rate
. Even though the logger will not
log data every time an event is checked, it will take a reading and hence will use the same
amount of battery power as a logged data point. Always choose the longest logging / event
checking interval that you can for your application in order to prolong the battery life.
1.5.
LeveLine
®
Factory Calibration
All LeveLine
®
water level Loggers are factory calibrated at five separate pressure points
across their pressure range, at two temperatures, 5°C and 25°C. In addition, the -CTD
versions are also calibrated at a single electrical conductivity point.
All LeveLine
®
Baro Loggers are factory calibrated at three separate pressure points across
their pressure range, at two temperatures, 5°C and 25°C.
The data produced by this calibration procedure is used to construct a three dimensional
calibration curve (surface), which is stored in each Probe's memory and subsequently used
to automatically correct temperature and pressure readings prior to output.
The date at which your LeveLine
®
Logger was last calibrated is shown in the 'LeveLine
Status' box on the Setup tab of the LeveLink
™
PC software.
1.5.1. Factory Re-calibration of Level and Temperature
The accuracy of LeveLine
®
Loggers can be adversely affected by physical damage or
abuse, improper care and handling, exceeding operating pressure and temperature limits,
lightning strikes and similar power surges, as well as normal drift in the device’s electronic
components.
Excluding physical damage, the need for factory recalibration is dependent upon the
amount of drift you are willing to tolerate.