You will be prompted to confirm that you want to calibrate EC and to confirm that the
correct calibration standard has been entered.
If the software detects a problem, an error message will be displayed. Possible errors are:
•
Temperature out of range. Must be between 5ºC and 40ºC (41ºF -104ºF).
•
EC Calibration value out of range. Must be between 100µS/cm and 99,999µS/cm.
•
Large differential between the selected calibration value and the actual EC reading,
which could indicate an incorrect calibration solution value
•
Readings are unstable.
If any of these errors are displayed, take remedial action and try again.
If calibration is successful, a 'Calibration Complete' message will be displayed and the
dedicated EC calibration boxes will be refreshed with new values.
After calibration, rinse the Probe in fresh water.
2.7.4. EC Cell Constant
The EC cell constant is an indication of the health of the EC cell. Nominally, the EC cell
constant should be around 1.25. This can vary with different calibration standard values
and calibration temperatures.
If the EC cell constant is reported above 2.00 or below 0.5, carefully clean the EC cell
contacts and the inside of the nose sleeve. Also check your calibration solution is fresh and
that you have input the quoted value (at 25ºC) during calibration.
2.7.5. Frequency of EC Calibration
Ideally, EC calibration should be carried out regularly based upon the levels of soiling the
Probe experiences.
On a regular basis, thoroughly clean the inside of the nose sleeve and the four metal EC
contacts situated in the EC cell with a soft cloth or toothbrush and non-abrasive detergent.
Never use solvent or alcohol based products to clean the EC cell
. After cleaning,
replace the nose sleeve and re-calibrate.
2.8. LeveLine
®
Setup In Detail
The following sections provide detailed guidance for setting up a LeveLine
®
Logger and
assumes you have connected a LeveLine
®
Logger to your PC and connected to it by
clicking the 'Connect to LeveLine' button.
Once a LeveLine
®
is connected, the 'Disconnect LeveLine' button will become live and
should be used prior to disconnecting the LeveLine
®
unit.