6.5.2. Address Query
This command allows the data recorder to obtain a sensor's device address. It must only be
issued when there is just one sensor attached to the bus, otherwise all sensors will
respond, causing bus contention.
Command Response
?!
<a><CR><LF>
6.5.3. Change Address
This command allows the data recorder to assign a new device address to the LeveLine
®
.
Command Response
<a>A<b>! <b><CR><LF>
•
<b> is the new device address to be assigned to the unit (1 character, value “0”-“9”,
“A”-“Z” or “a”-“z”).
6.5.4. Send Identification
This command allows the data recorder to retrieve some basic information about the
LeveLine
®
.
Command Response
<a>I!
<a><sdi-ver><vendor-id><model-id><firmware-
ver><serial-nr>
<CR><LF>
•
<sdi-ver> is the version of the supported SDI-12 standard (“13” for version 1.3).
•
<vendor-id> is an 8-character vendor identification string (“AQUAREAD”).
•
<model-id> is a 6-character device identification string (see note below).
•
<firmware-ver> is a 3-digit LeveLine
®
firmware version string (e.g. “100” for v1.00).
•
<serial-nr> is the LeveLine's serial number string (8 characters).
Model Identification Strings
•
For the standard LeveLine
®
Probe the <model id> is (“LVLINE”).
•
For the LeveLine - CTD
®
Probe the <model id> is (“LL-CTD”).
•
For the standard LeveLine
®
Mini Probe the <model id> is (“LLMINI”).
•
For the LeveLine
®
Mini - CTD Probe the <model id> is (“LLMCTD”).
6.5.5. Start Measurement
This command tells the LeveLine
®
to start making a measurement.
Command
Response
<a>M!
<a><time><nr-values><CR><LF>
•
<time> is the time in seconds (three digits) until the measurement will be ready. The
LeveLine
®
always places a value of “001” in this field.
•
<nr-values> is a single digit specifying the number of measurements that will be
returned by subsequent Send Data commands. The LeveLine
®
always places a
value of “2” in this field.