LeveLine
®
Instruction Manual
10401-01026 Rev D
Sequence initiated with Start Measurement command (-CTD Probe):
Command
Response
<a>M!
or
<a>MC!
<a>0016<CR><LF>
<a>D0!
<a><temperature><pressure><ABS-EC><EC-20><CR><LF>
or
<a><temperature><pressure><ABS-EC><EC-20><crc><CR><LF>
<a>D1!
<a><EC-25><SAL><CR><LF>
or
<a><EC-25><SAL><crc><CR><LF>
Sequence initiated with Start Concurrent Measurement command (-CTD Probe):
Command
Response
<a>C!
or
<a>CC!
<a>00106<CR><LF>
<a>D0!
<a><temperature><pressure><ABS-EC><EC-20><EC-25><SAL>
<CR><LF>
or
<a><temperature><pressure><ABS-EC><EC-20><EC-25><SAL>
<crc><CR><LF>
Where:
•
ABS-EC is the absolute (uncorrected) EC value in µS/cm.
•
EC-20 is EC in µS/cm corrected to 20°C.
•
EC-25 is EC in µS/cm corrected to 25°C.
•
SAL is salinity in PSU, which is an equivalent to ppt.
In the event of an error or over range condition, EC or SAL values will be sent as negative.