Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
29
USB 2.0 Triggering and Serial Decode
6
For both the D+ and D- signals (for low or full speed, similar steps for
the single Source for high speed):
a
Connect an oscilloscope channel to the signal in the device under
test.
b
Press the
D+ Source
or
D- Source
softkey; then, turn the Entry knob to
select the channel for the signal.
c
Press the corresponding
Threshold
softkey; then, turn the Entry knob
to select the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
7
Press
Auto Setup
to automatically set these options for decoding and
triggering on USB signals:
•
Low- speed:
•
D+/- Source thresholds: 1.4 V
•
D+/- Source vertical scale: 1.0 V/div
•
D+/- Source vertical offset: 0.0 V
•
Horizontal scale: 5 µs/div
•
Full- speed:
•
D+/- Source thresholds: 1.4 V
•
D+/- Source vertical scale: 1.0 V/div
•
D+/- Source vertical offset: 0.0 V
•
Horizontal scale: 500 ns/div
•
High- speed:
•
D+/- Source thresholds: 0.0 V
•
D+/- Source vertical scale: 200 mV/div
•
D+/- Source vertical offset: 0.0 V
•
Horizontal scale: 20 ns/div
•
Serial Decode: On
•
Trigger Mode: currently active serial bus
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...