Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
27
MIL-STD-1553/ARINC 429 Triggering and Serial Decode
•
Sync Error
— finds invalid Sync pulses.
•
Manchester Error
— finds Manchester encoding errors.
For more information on searching data, see
"Searching Lister Data"
on
page 146.
For more information on using the
[Navigate]
key and controls, see
"Navigating the Time Base"
on page 75.
Setup for ARINC 429 Signals
Setup consists of first connecting the oscilloscope to a ARINC 429 signal
using a differential active probe (the Agilent N2791A is recommended),
then using the Signals Menu to specify the signal source, the high and low
trigger threshold voltage levels, the signal speed, and the signal type.
To set up the oscilloscope to capture ARINC 429 signals:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Serial
softkey, turn the Entry knob to select the desired slot
(Serial 1 or Serial 2), and press the softkey again to enable decode.
4
Press the
Mode
softkey; then, select
ARINC 429
decode mode.
5
Press the
Signals
softkey to open the ARINC 429 Signals Menu.
6
Press
Source
; then, select the channel for the ARINC 429 signal.
The label for the ARINC 429 source channel is automatically set.
7
Press the
Speed
softkey and specify the speed of the ARINC 429 signal:
•
High
— 100 kb/s.
•
Low
— 12.5 kb/s.
8
Press the
Signal Type
softkey and specify the signal type of the ARINC
429 signal:
•
Line A (non-inverted)
.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...