Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
11
Trigger Mode/Coupling
See Also
•
"Forcing a Trigger"
on page 167
•
"To set the trigger holdoff"
on page 210
•
"To position the time reference (left, center, right)"
on page 73
To select the trigger coupling
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Coupling
softkey;
then, turn the Entry knob to select:
•
DC
coupling — allows DC and AC signals into the trigger path.
•
AC
coupling — places a 10 Hz high- pass filter in the trigger path
removing any DC offset voltage from the trigger waveform.
The high- pass filter in the External Trigger input path is 50 Hz for
all models.
Use AC coupling to get a stable edge trigger when your waveform has
a large DC offset.
•
LF
(low frequency)
Reject
coupling — adds a high- pass filter with the
3- dB point at 50 kHz in series with the trigger waveform.
Low frequency reject removes any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
etc., that can interfere with proper triggering.
Use
LF Reject
coupling to get a stable edge trigger when your
waveform has low frequency noise.
Pass
Band
50 kHz
0 dB
3 dB down point
DC
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...