I2S Triggering and Serial Decode
26
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
439
3
Press the
Trigger Setup
softkey to open the I
2
S Trigger Setup Menu.
4
Press the
Audio
softkey; then, turn the Entry knob to choose to trigger
on
Left
channel events,
Right
channel events, or events that occur on
Either
channel.
5
Press the
Trigger
softkey and choose a qualifier:
•
Equal
— triggers on the specified audio channel's data word when it
equals the specified word.
•
Not equal
— triggers on any word other than the specified word.
•
Less than
— triggers when the channel's data word is less than the
specified value.
•
Greater than
— triggers when the channel's data word is greater than
the specified value.
•
In Range
— enter upper and lower values to specify the range in
which to trigger.
•
Out of Range
— enter upper and lower values to specify range in which
trigger will not occur.
•
Increasing value
— triggers when the data value is increasing over time
and the specified value is met or exceeded. Set
Trigger >=
to the data
value that must be reached. Set
Armed <=
to the value to which the
data must fall before the trigger circuit is re- armed (ready to trigger
again). These settings are made in the current menu when
Base
is
Decimal
or in the Bits submenu when the
Base
is
Binary
. The Armed
control reduces triggers due to noise.
This trigger condition is best understood when the digital data
transferred over the I2S bus is considered in terms of representing
an analog waveform. The figure below shows a plot of sample data
transmitted over an I2S bus for one channel. In this example, the
oscilloscope will trigger at the 2 points shown since there are two
instances in which the data increases from a value below (or at) the
"Armed" value to a value greater than (or equal to) the specified
"Trigger" value.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...