464
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
28
UART/RS232 Triggering and Serial Decode
•
Rx Stop Bit
— Triggers when a stop bit occurs on Rx. The trigger will
occur on the first stop bit. This is done automatically whether the
device under test uses 1, 1.5, or 2 stop bits. You do not need to
specify the number of stop bits used by the device Under test.
•
Rx Data
— Triggers on a data byte that you specify. For use when the
device under test data words are from 5 to 8 bits in length (no 9th
(alert) bit).
•
Rx 1:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 1. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
•
Rx 0:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 0. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
•
Rx X:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers on a data byte
that you specify regardless of the value of the 9th (alert) bit. The
specified data byte applies to the least significant 8 bits (excludes
the 9th (alert) bit).
•
Similar choices are available for Tx.
•
Rx or Tx Parity Error
— Triggers on a parity error based on the parity
you have set in the Bus Configuration Menu.
6
If you choose a trigger condition that includes "
Data
" in its description
(for example:
Rx Data
), then press the
Data is
softkey, and choose an
equality qualifier. You can choose equal to, not equal to, less than, or
greater than a specific data value.
7
Use the
Data
softkey to choose the data value for your trigger
comparison. This works in conjunction with the
Data is
softkey.
8
Optional: The
Burst
softkey lets you trigger on the Nth frame (1- 4096)
after an idle time you select. All trigger conditions must be met for the
trigger to occur.
9
If
Burst
is selected, an idle time (1 µs to 10 s) can be specified so that
the oscilloscope will look for a trigger condition only after the idle time
has past. Press the
Idle
softkey and rotate the Entry knob to set an idle
time.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...