220
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
•
Normal
— at slower time/div settings, normal decimation occurs, and
there is no averaging. Use this mode for most waveforms. See
"Normal Acquisition Mode"
on page 220.
•
Peak Detect
— at slower time/div settings, the maximum and minimum
samples in the effective sample period are stored. Use this mode for
displaying narrow pulses that occur infrequently. See
"Peak Detect
Acquisition Mode"
on page 220.
•
Averaging
— at all time/div settings, the specified number of triggers
are averaged together. Use this mode for reducing noise and
increasing resolution of periodic signals without bandwidth or rise
time degradation. See
"Averaging Acquisition Mode"
on page 223.
•
High Resolution
— at slower time/div settings, all samples in the
effective sample period are averaged and the average value is stored.
Use this mode for reducing random noise. See
"High Resolution
Acquisition Mode"
on page 225.
Normal Acquisition Mode
In Normal mode at slower time/div settings, extra samples are decimated
(in other words, some are thrown away). This mode yields the best display
for most waveforms.
Peak Detect Acquisition Mode
In Peak Detect mode at slower time/div settings, minimum and maximum
samples are kept in order to capture infrequent and narrow events (at the
expense of exaggerating any noise). This mode displays all pulses that are
at least as wide as the sample period.
For InfiniiVision 4000 X- Series oscilloscopes, which have a maximum
sample rate of 5 GSa/s, a sample is taken every 200 ps (sample period).
See Also
•
"Glitch or Narrow Pulse Capture"
on page 221
•
"Using Peak Detect Mode to Find a Glitch"
on page 222
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...