26
I2S Triggering and Serial Decode
If you select an "Armed" value that is equal to or greater than the
"Trigger" value, the "Trigger" value will be increased so that it is
always greater than the "Armed" value.
•
Decreasing value
— similar to the description above except the trigger
occurs on a decreasing data word value, and the "Armed" value is
the value to which the data must rise in order to re- arm the trigger.
6
Press the
Base
softkey and select a number base for entering data
values:
•
Binary (2's complement)
.
When Binary is selected, the
Bits
softkey appears. This softkey opens
the I2S Bits Menu for entering data values.
When the trigger qualifier requires a pair of values (as with In
Range, Out of Range, Increasing value, or Decreasing value), the first
softkey in the I2S Bits Menu lets you select which value of the pair.
In the I2S Bits Menu, press the
Bit
softkey and rotate the Entry knob
to select each bit; then, use the
0 1 X
softkey to set each bit value to
zero, one, or don't care. You can use the
Set all Bits
softkey to set all
bits to the value chosen on the
0 1 X
softkey. Don't care values are
only allowed with Equal or Not Equal trigger qualifiers.
•
Signed decimal
.
When Decimal is selected, the softkey(s) to the right let you enter
decimal values with the Entry knob. These softkeys can be
Data
,
<
,
>
,
or
Threshold
depending on the selected trigger qualifier.
7ULJJHUYDOXH
7ULJJHU
7ULJJHU
1RWULJJHU
$UPHGYDOXH
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...