Setup for LIN Signals
LIN (Local Interconnect Network) signal setup consists of connecting the
oscilloscope to a serial LIN signal, specifying the signal source, threshold
voltage level, baud rate, sample point, and other LIN signal parameters.
To set up the oscilloscope to capture LIN signals:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Serial
softkey, turn the Entry knob to select the desired slot
(Serial 1 or Serial 2), and press the softkey again to enable decode.
4
Press the
Mode
softkey; then, select
LIN
trigger type.
5
Press the
Signals
softkey to open the LIN Signals Menu.
6
Press the
Source
softkey to select the channel connected to the LIN
signal line.
The label for the LIN source channel is automatically set.
7
Press the
Threshold
softkey; then, turn the Entry knob to set the LIN
signal threshold voltage level to the middle of the LIN signal.
The threshold voltage level is used in decoding, and it will become the
trigger level when the trigger type is set to the selected serial decode
slot.
8
Press the
Baud Rate
softkey to open the LIN Baud Rate Menu.
9
Press the
Baud
softkey; then, turn the Entry knob to select the baud
rate that matches match your LIN bus signal.
The default baud rate is 19.2 kb/s.
If none of the pre- defined selections match your LIN bus signal, select
User Defined
; then, press the
User Baud
softkey and turn the Entry knob
to enter the baud rate.
You can set the LIN baud rate from 2.4 kb/s to 625 kb/s in increments
of 100 b/s.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...