Utility Settings
20
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
339
The real- time clock only allows selection of valid dates. If a day is
selected and the month or year is changed so the day is invalid, the day
is automatically adjusted.
Setting the Rear Panel TRIG OUT Source
You can choose the source of the TRIG OUT connector on the rear panel
of the oscilloscope:
1
Press
[Utility] > Options > Rear Panel
.
2
In the Rear Panel Menu, press
Trig Out
; then, turn the Entry knob to
select from:
•
Triggers
— Each time the oscilloscope triggers, a rising edge occurs on
TRIG OUT. The rising edge is delayed 30 ns from the oscilloscope's
trigger point. The output level is 0- 5 V into an open circuit, and
0- 2.5 V into 50
Ω
. See
Chapter 10
, “Triggers,” starting on page 165.
•
Mask
— The pass/fail status is evaluated periodically. When the
evaluation of the testing period results in a failure, the trigger output
pulses high (+5 V). Otherwise, the trigger output remains at low
(0 V). See
Chapter 15
, “Mask Testing,” starting on page 271.
•
Waveform Generator 1/2 Sync Pulse
— All of the waveform generator
output functions (except DC, Noise, and Cardiac) have an associated
Sync signal:
The Sync signal is a TTL positive pulse that occurs when the
waveform rises above zero volts (or the DC offset value).
See
Chapter 17
, “Waveform Generator,” starting on page 287.
The TRIG OUT connector also provides the User Cal signal. See
"To
perform user calibration"
on page 342.
Setting the Reference Signal Mode
The
10 MHz REF
BNC connector on the rear panel is provided so you
can:
•
Supply a more accurate sample clock signal to the oscilloscope, or
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...