Measurements
14
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
269
You can press the
Display On
softkey again to disable the measurement
statistics display. Statistics continue to accumulate even when the
statistics display is disabled.
5
To reset the statistics measurements, press the
Reset Statistics
softkey.
This resets all statistics and begins recording statistical data again.
Each time a new measurement (for example: frequency, period, or
amplitude) is added the statistics are reset and accumulation of
statistical data begins again.
6
To enable a relative standard deviation, press the
Relative
σ
softkey.
When enabled, the standard deviation shown in measurement statistics
becomes the standard deviation/mean.
7
To specify the number of values used when calculating measurement
statistics, press the
Max Count
softkey and enter the desired value.
Other things to know about measurement statistics:
•
When the
[Single]
key is pressed, statistics are reset and a single
measurement is done (count = 1). Successive
[Single]
acquisitions
accumulate statistical data (and the count is incremented).
•
The
Increment Statistics
softkey only appears when the acquisition is
stopped and the optional segmented memory feature is off. Press the
[Single]
or
[Run/Stop]
key to stop the acquisition. You can use the
horizontal position control (in the Horizontal control section of the
front panel) to pan through the waveform. Active measurements will
stay on screen, allowing you to measure various aspects of the captured
waveforms. Press
Increment Statistics
to add the currently measured
waveform to the collected statistical data.
•
The
Analyze Segments
softkey only appears when the acquisition is
stopped and the optional segmented memory feature is on. After an
acquisition has completed (and the oscilloscope is stopped), you can
press the
Analyze Segments
softkey to accumulate measurement statistics
for the acquired segments.
You can also turn on infinite persistence (in the Display Menu) and
press the
Analyze Segments
softkey to create an infinite persistence
display.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...