Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
25
I2C/SPI Triggering and Serial Decode
•
If the framing signal is set to
Timeout
, the oscilloscope generates
it's own internal framing signal after it sees inactivity on the serial
clock line.
Clock Timeout
— Select
Clock Timeout
in the
Frame by
softkey, then
select the
Timeout
softkey and turn the Entry knob to set the
minimum time that the Clock signal must be idle (not
transitioning) before the oscilloscope will search for the Data
pattern on which to trigger.
The Timeout value can be set anywhere from 100 ns to 10 s.
When you press the
Frame by
softkey, the
Display Info
graphic changes
to show timeout selection or the current state of the chip select
signal.
b
Press the
Threshold
softkey; then, turn the Entry knob to select the
chip select signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
When
Display Info
is enabled, information about the selected signal sources
and their threshold voltage levels, as well as a waveform diagram, appears
on the screen.
SPI Triggering
To set up the oscilloscope to capture SPI signals, see
"Setup for SPI
Signals"
on page 424.
After the oscilloscope has been set up to capture SPI signals, you can then
trigger on a data pattern that occurs at the start of a frame. The serial
data string can be specified to be from 4 to 32 bits long.
When you select the SPI trigger type and
Display Info
is enabled, a graphic
is displayed showing the current state of the frame signal, clock slope,
number of data bits, and data bit values.
1
Press
[Trigger]
.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...