6
Press
Source
and select the analog channel that is probing the FlexRay
signal.
7
Press
Threshold
; then, turn the Entry knob to set the threshold voltage
level.
The threshold level should be set below the idle level.
The threshold voltage level is used in decoding and will become the
trigger level when the trigger type is set to the selected serial decode
bus.
8
Press
Baud
and select the baud rate of the FlexRay signal being probed.
9
Press
Bus
and select the bus type of the FlexRay signal being probed.
It is important to specify the correct bus because this setting affects
CRC error detection.
10
Press
Auto Setup
to perform the following actions:
•
Sets the selected source channel's impedance to 50 Ohms, assuming a
differential active probe that requires a 50 ohm termination is being
used .
•
Sets the selected source channel's probe attenuation to 10:1.
•
Sets the trigger level (on the selected source channel) to - 300 mV.
•
Turns on trigger Noise Reject.
•
Turns on Serial Decode.
•
Sets the trigger type to FlexRay.
FlexRay Triggering
To set up the oscilloscope to capture a FlexRay signal, see
"Setup for
FlexRay Signals"
on page 405.
After you have set up the oscilloscope to capture a FlexRay signal, you
can then set up triggers on frames (see
page 407
), errors (see
page 408
),
or events (see
page 408
).
N O T E
To display FlexRay serial decode, see
"FlexRay Serial Decode"
on page 409.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...