Measurements
14
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
265
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an
integral number of periods of the displayed signal. If less than three edges
are present, the measurement shows "No edges".
The X cursors show what interval of the waveform is being measured.
Measurement Thresholds
Setting measurement thresholds defines the vertical levels where
measurements will be taken on an analog channel or math waveform.
1
From the Measurement Menu, press the
Settings
softkey; then, press the
Thresholds
softkey to set analog channel measurement thresholds.
You can also open the Measurement Threshold Menu by pressing
[Analyze] > Features
and then selecting
Measurement Thresholds
.
2
Press the
Source
softkey to select the analog channel or math waveform
source for which you want to change measurement thresholds.
T
Negative area
N cycles
Positive area
N O T E
Changing default thresholds may change measurement results
The default lower, middle, and upper threshold values are 10%, 50%, and 90% of the value
between Top and Base. Changing these threshold definitions from the default values may
change the returned measurement results for Average, Delay, Duty Cycle, Fall Time,
Frequency, Overshoot, Period, Phase, Preshoot, Rise Time, +Width, and -Width.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...