Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
22
Reference
Waveform Data
Header
A waveform may have more than one data set. Each waveform data set
will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.
X Units
A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
• 0 = Unknown.
• 1 = Volts.
• 2 = Seconds.
• 3 = Constant.
• 4 = Amps.
• 5 = dB.
• 6 = Hz.
Y Units
A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
Date
A 16-byte character array, left blank in InfiniiVision oscilloscopes.
Time
A 16-byte character array, left blank in the InfiniiVision oscilloscopes.
Frame
A 24 byte character array that is the model number and serial number of the
oscilloscope in the format of: MODEL#:SERIAL#.
Waveform Label
A 16 byte character array that contains the label assigned to the waveform.
Time Tags
A 64-bit double, only used when saving multiple segments (requires
segmented memory option). This is the time (in seconds) since the first
trigger.
Segment Index
A 32-bit unsigned integer. This is the segment number. Only used when saving
multiple segments.
Waveform Data
Header Size
A 32-bit integer that is the size of the waveform data header.
Buffer Type
A 16-bit short that is the type of waveform data stored in the file:
• 0 = Unknown data.
• 1 = Normal 32-bit float data.
• 2 = Maximum float data.
• 3 = Minimum float data.
• 4 = Not used in InfiniiVision oscilloscopes.
• 5 = Not used in InfiniiVision oscilloscopes.
• 6 = Digital unsigned 8-bit char data (for digital channels).
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...