Acquisition Control
The
Analyze Segments
softkey appears when the acquisition is stopped and
the segmented memory feature is on or when the serial Lister is enabled.
You can also turn on infinite persistence (in the Display Menu) and press
the
Analyze Segments
softkey to create an infinite persistence display.
Segmented Memory Re-Arm Time
After each segment fills, the oscilloscope re- arms and is ready to trigger
in about 1 µs.
Remember though, for example: if the horizontal time per division control
is set to 5 µs/div, and the Time Reference is set to
Center
, it will take at
least 50 µs to fill all ten divisions and re- arm. (That is 25 µs to capture
pre- trigger data and 25 µs to capture post- trigger data.)
Saving Data from Segmented Memory
You can save either the currently displayed segment (
Save Segment -
Current
), or all segments (
Save Segment - All
) in the following data formats:
CSV, ASCII XY, and BIN.
Be sure to set the Length control to capture enough points to accurately
represent the captured data. When the oscilloscope is busy saving multiple
segments, progress is displayed in the upper right area of the display.
For more information, see
"To save CSV, ASCII XY, or BIN data files"
on
page 312.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...