I2C/SPI Triggering and Serial Decode
25
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
425
To set up the oscilloscope to capture SPI signals, use the
Signals
softkey
which appears in the Serial Decode Menu:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Serial
softkey, turn the Entry knob to select the desired slot
(Serial 1 or Serial 2), and press the softkey again to enable decode.
4
Press the
Mode
softkey; then, select
SPI
trigger type.
5
Press the
Signals
softkey to open the SPI Signals Menu.
6
Press the
Clock
softkey to open the SPI Clock Menu.
In the SPI Clock Menu:
a
Press the
Clock
softkey; then, turn the Entry knob to select the
channel connected to the SPI serial clock line.
The CLK label for the source channel is automatically set.
b
Press the
Threshold
softkey; then, turn the Entry knob to select the
clock signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
c
Press the slope softkey (
) to select rising edge or falling edge for
the selected Clock source.
This determines which clock edge the oscilloscope will use to latch
the serial data. When
Display Info
is enabled, the graphic changes to
show the current state of the clock signal.
7
Press the
MOSI
softkey to open the SPI Master- Out Slave- In Menu.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...