Trigger Mode/Coupling
11
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
207
In either Auto or Normal trigger mode, a trigger may be missed if the
event occurs while the pre- trigger buffer is being filled. This may be more
likely, for example, when the horizontal scale knob is set to a slow
time/div setting, such as 500 ms/div.
Trigger Indicator
The trigger indicator at the top right of the display shows whether triggers
are occurring.
In the
Auto
trigger mode, the trigger indicator can show:
•
Auto?
— the trigger condition is not found (after the pre- trigger buffer
has filled), and forced triggers and acquisitions are occurring.
•
Auto
— the trigger condition is found (or the pre- trigger buffer is being
filled).
In the
Normal
trigger mode, the trigger indicator can show:
•
Trig'd?
— the trigger condition is not found (after the pre- trigger buffer
has filled), and no acquisitions are occurring.
•
Trig'd
— trigger condition is found (or pre- trigger buffer is being filled).
When the oscilloscope is not running, the trigger indicator area shows
Stop
.
When to Use Auto
Trigger Mode
The
Auto
trigger mode is appropriate when:
•
Checking DC signals or signals with unknown levels or activity.
•
When trigger conditions occur often enough that forced triggers are
unnecessary.
When to Use
Normal Trigger
Mode
The
Normal
trigger mode is appropriate when:
•
You only want to acquire specific events specified by the trigger
settings.
•
Triggering on an infrequent signal from a serial bus (for example, I2C,
SPI, CAN, LIN, etc.) or another signal that arrives in bursts. The
Normal
trigger mode lets you stabilize the display by preventing the oscilloscope
from auto- triggering.
•
Making single- shot acquisitions with the
[Single]
key.
Often with single- shot acquisitions, you must initiate some action in the
device under test, and you don't want the oscilloscope to auto- trigger
before that happens. Before initiating the action in the circuit, wait for
the trigger condition indicator
Trig'd?
to display (this tells you the
pre- trigger buffer is filled).
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...