Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
Acquiring to Segmented Memory
When capturing multiple infrequent trigger events it is advantageous to
divide the oscilloscope's memory into segments. This lets you capture
signal activity without capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see
"Measurements, Statistics, and Infinite Persistence with Segmented
Memory"
on page 229) to show infinite persistence across all acquired
segments. See also
"To set or clear persistence"
on page 151 for details.
To acquire to
segmented
memory
1
Set up a trigger condition. (See
Chapter 10
, “Triggers,” starting on page
165 for details.)
2
Press the
[Acquire]
key in the Waveform section of the front panel.
3
Press the
Segmented
softkey.
4
In the Segmented Memory Menu, press the
Segmented
softkey to enable
segmented memory acquisitions.
5
Press the
# of Segs
softkey and turn the Entry knob to select the
number of segments into which you would like to divide the
oscilloscope's memory.
Memory can be divided into as few as two segments and as many as
1000 segments, depending on the oscilloscope model.
6
Press the
[Run]
or
[Single]
key.
The oscilloscope runs and fills a memory segment for each trigger event.
When the oscilloscope is busy acquiring multiple segments, the progress is
displayed on screen. The oscilloscope continues to trigger until memory is
filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity,
consider selecting the
Normal
trigger mode to prevent AutoTriggering. See
"To select the Auto or Normal trigger mode"
on page 206.
See Also
•
"Navigating Segments"
on page 229
•
"Measurements, Statistics, and Infinite Persistence with Segmented
Memory"
on page 229
•
"Segmented Memory Re- Arm Time"
on page 230
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...