![Agilent Technologies DSO-X 4022A User Manual Download Page 256](http://html1.mh-extra.com/html/agilent-technologies/dso-x-4022a/dso-x-4022a_user-manual_2868638256.webp)
256
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
14
Measurements
Period
Period is the time period of the complete waveform cycle. The time is
measured between the middle threshold points of two consecutive,
like- polarity edges. A middle threshold crossing must also travel through
the lower and upper threshold levels which eliminates runt pulses. The X
cursors show what portion of the waveform is being measured. The Y
cursor shows the middle threshold point.
Frequency
Frequency is defined as 1/Period. Period is defined as the time between
the middle threshold crossings of two consecutive, like- polarity edges. A
middle threshold crossing must also travel through the lower and upper
threshold levels which eliminates runt pulses. The X cursors show what
portion of the waveform is being measured. The Y cursor shows the
middle threshold point.
See Also
•
"To isolate an event for frequency measurement"
on page 256
To isolate an event for frequency measurement
The following figure shows how to use Zoom mode to isolate an event for
a frequency measurement.
You may need to change the measurement window setting so that the
measurement is made in the lower, Zoom window. See
"Measurement
Window"
on page 267.
If the waveform is clipped, it may not be possible to make the
measurement.
Summary of Contents for DSO-X 4022A
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...