Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 66 of 84
f. Remove the fixed 100µm aperture and set it aside.
g. Move distance marker 1 to the top of the furthest tic mark on the left-
side of the cropped window (tic mark 30 in the illustration).
Move
distance marker 2 to the top of the farthest tic mark on the right-side of the
cropped window (tic mark 70 in the illustration).
h. Count the distance between the markers.
Each tic mark on the calibration
scale is 10µm, or 100µm between two medium or two long tic marks. Enter
the distance in the
Distance bet ween points
dialog box; Change the
Objective magnification
to 10X and enter 10X in the
brief and full description
dialog boxes. Click Save to exit the window.
400µ