Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 50 of 84
g. While viewing the interferogram, change the Gain DAC 1X as
appropriate to increase the interferogram MAX peak to ~7 volts.
h. Lower the condenser stage.
i. Close BDiag.
30. Tune the sluice (reflection) mirror.
a. Switch the iN5 to reflection mode by moving the knob on the left side of
the iN5.
Reflection mode