Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 54 of 84
37. Remove the 100µm fixed sample aperture and set aside.
38. Reinstall the X-Y manual stage and sample holder.
39. Adjust the binocular viewer.
If a camera only is installed, continue with step 40.
a. Switch the microscope to reflection mode.
Reflection mode
b. Install the calibration slide into the sample holder so the text on the
slide is upside down when installed.
c. While looking through the eyepieces move the X-Y stage to find the
calibration scale and center the scale in the field of view.