Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 64 of 84
i. If you have a camera only installed, adjust the z stage to focus on
the calibration scale.
Then move the X-Y stage to center the edge of
the 50µm tic mark so it is positioned at the exact center of the gray
crosshair in the center of the video calibration window.
c. Insert the 100µm fixed aperture.
The aperture should be installed with the
text properly orientated and facing out (toward you).
Note
If you have a binocular viewer and camera installed, ensure the
aperture image is centered on the reticle as viewed through the right
eyepiece otherwise
repeat step 27 before proceeding with step 44d.