Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 47 of 84
Note
Only the default files that correspond to the detector installed will be
available on a new system.
29. Tune the input mirror.
a. Select the bench tab and ensure the Sample Compartment = Right
µScope, % T.
Leave the window OPEN.
b. If you have an iS10, tune the condenser for best IR throughput.
This is
because of the location of the source relative to the aperture.
If you have an iS50, continue with step 29c.