Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 20 of 84
i. Secure the interface.
Using a 5/32 inch balldriver and the mounting screws
set aside in step 9c,
loosely
thread the 2 screws into the locations shown
below and
then
tighten the screws to secure the interface to the iS50.
Install short screw in sample compartment
Install long screw through the passport mirror baseplate
j. Reinstall the iS50 main cover, sample compartment plate and sample
compartment cover by reversing steps 8a-8d and then continue with
step 10.