Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 52 of 84
c. Using a 5/64 inch balldriver, adjust the two opposing screws on the
sluice mirror for best MAX interferogram peak.
MCT MAX peak should be ~1/2 the throughput of transmission mode
DTGS MAX peak should be ~1/2 the throughput of transmission mode
Note
When tuning the sluice mirror, the mirror should be manually switched
between %T and %R mode to ensure the sluice mirror is seated on the hard-
stop between adjustments.