Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 30 of 84
iii. Ensure the hole and the tip of the screw line up and then slide
the condenser towards the thumbscrew until it stops and hold
the condenser securely.
Then, gently push the thumbscrew into
threaded hole to engage the condenser and then snug the screw so
the condenser is secure. You should be able to
slightly
move the
condenser back and forth enough to adjust the X-Y alignment screws.
Note
DO NOT cross thread the screw or damage to the screw and
condenser may occur.
Note
DO NOT fully compress the spring or X-Y adjustments will not
be possible and signal may be compromised.
Condenser properly installed and secure
Do NOT fully compress
the spring