Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 31 of 84
f. Install the X-Y stage.
i. Using the z stage focus knobs, turn the knobs clockwise (away
from you) to lower the z stage so the z stage dovemount is just
above the condenser.
ii. Install the manual X-Y stage.
Slightly tilt the stage “up” at the front;
engage the tabs
under the back-end of stage circular opening with
the
back edge of the z stage dovemount and then lower the front end of
the stage.
Tilt stage back to engage tabs Stage inverted to show tabs
Dovemount slightly
above condenser