Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 69 of 84
f. Select the Collect tab, change the following parameters and click OK.
g. From the Collect drop-down menu select collect background and then
click Start Collection.
You can choose either yes or no to add the
background to the current window.
h. Using a 5/64 inch balldriver tighten the 2-plungers screws in the slider
and then back-off approximately ¾ of a turn.
i. Lower the z stage to make clearance for the tip ATR and then insert the
ATR all the way into the objective until it “clicks” into place.