Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 71 of 84
m. Start the sample collection to complete the rotational alignment and
save the results if applicable.
n. Remove the sample holder with slide from the X-Y stage and set aside.
o. Remove the 100µm fixed aperture and set aside.
p. Lower the z stage to make clearance for the tip ATR and pull the slide
out half-way until the ATR “clicks” into place (view mode).
q. If you have a camera only option, open Microview if not already open;
otherwise use the binocular viewer to perform the following ATR X-Y
alignment as it provides a larger field of view.
r. Insert a polycarbonate slide on the contact alert plate and tape down on
the edges to secure the slide.