Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 48 of 84
c. Using a 5/64 inch balldriver, adjust the two opposing screws on the
input mirror for best MAX interferogram peak.
MCT should be ~7 volts MAX
DTGS should be 2.5 volts MAX or greater
d. If the Max peak is between 7-7.5 volts continue with step 30 otherwise
continue with step 29e.
e. Open BDiag>Smart Devices>Smart Detector Preamp Boards or
Detector Preamps window depending on if you have an iS50 or iS10
respectively.
If you have an iS10 spectrometer; select iN5 and click OK.