Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 23 of 84
d. Using a 5/16 inch T-handle or L-wrench turn the screw in the interface
clockwise until it stops.
The latch should engage with the locator pin under
the iN5 and lock it into position. If not, reset the latch and then reconnect.
Note
When first turning the screw the iN5 may appear to push slightly away
from the interface. Hold the iN5 secure and continue to turn the screw until
the latch pulls the iN5 toward the interface. Once the iN5 touches the
interface, continue to turn the T-wrench approximately 1/4 turn until it stops.
If you use an L-wrench you will need to use the short side for leverage.
e. If you have an iS50, connect the touch point cable to the back of the
iS50 otherwise continue with step 10f.