Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 83 of 84
j. Select the Nicolet iN5 Microscope MCT-A or DTGS Detector KBr-EP test
and then click Qualify.
Follow the command prompts to complete the Transmission and
Reflection tests. When complete, remove the Validation plate and store
in the protective case.
IMPORTANT
You will need to manually move
and
focus the stage at each
background and sample location when prompted.
k. If you have an MCTA detector and Ge Tip ATR continue with step 58l
otherwise continue with step 59.
l. Install the contact alert plate on to the X-Y stage and then insert the 32
Mil Polystyrene sample slide on the plate.
m. Select the Ge Tip ATR Qualification test and follow the prompts to
complete the test.
Transmission Background
Center the open hole in the
beampath when ValPro asks
to move to the polystyrene
transmission location
Transmission Sample
Center and focus on the polystyrene
over the open hole when ValPro asks
to place polystyrene sample in the
sample holder
Reflection Background
Center and focus on the gold
mirror when ValPro asks to
insert gold slide
Reflection Sample
Center and focus on the gold mirror
with polystyrene when ValPro asks
to place polystyrene sample in the
sample holder
NOT USED on iN5