Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 67 of 84
i. From the brief description drop-down menu select 10X and click Set to
Current.
Click close to exit the window.
j. Remove the calibration slide from the sample holder.
45. Align the Ge Tip ATR; if you do not have a Ge Tip ATR, continue with step 46.
a. Install a gold or aluminum slide into the sample holder.
b. Using the z stage control knobs focus on the slide while looking at
Microview or through the binocular viewer.