Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 44 of 84
b. Focus condenser and adjust X-Y position.
Ensure the condenser is
focused by
fine tuning
the condenser knob if necessary using the dark
corner/corners as a reference.
If necessary, using two 5/64 inch balldrivers, move the condenser in the direction
necessary to view two dark corners if only one is visible. Then, adjust the
condenser X-Y alignment screws to center the circle of illumination using the
corners as a reference and then adjust the condenser straight up or down or
straight left or right until [all] corners are out of the field of view.
Out of focus
In focus