Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 81 of 84
56. Close diagnostics window.
57. Remove the sample holder and slide.
58. Run ValPro.
If you do not have ValPro, continue with step 59.
a. Switch the iN5 to transmission mode by moving the knob on the left-
side of the iN5.
Transmission mode
b. Remove the fixed 100µm sample aperture; raise the z stage and focus
and center the condenser.
c. Insert the fixed 100µm sample aperture.
Ensure the aperture image is
centered and focused as viewed through the eyepieces or the Microview
window.
d. Install the validation plate onto the X-Y stage.