n. Remove the 32 Mil sample slide and store in the protective case.
59. If a variable glass or black aperture were ordered, ensure the blades move
smoothly by turning the adjustment screws so the blades are fully open and
then fully closed. Install the variable aperture into the iN5 and demonstrate
use to the customer.
Note
A Full Field Aperture User Guide is included on iN5 software DVD. Refer the
customer to the guide for future reference.
THE INSTALLATION IS COMPLETE