Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 26 of 84
14. Prepare the detector.
a. If the iN5 is connected to an iS50 spectrometer, change the detector
switch settings (the iN5 is tested on an iS10 at the factory and the
detector switch settings are addressed for an iS10 spectrometer when
shipped).
DTGS MCT
b.
If the iN5 has an MCTA detector installed, wear the proper safety
equipment and use the thermos and funnel provided to fill the MCTA
detector with LN2 ensuring it is full (approximately 1 liter of LN2 when
warm).