Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 82 of 84
e. Using the X-Y stage knobs, move the stage so the open-hole
(transmission background) is centered under the objective.
Refer to the
illustration in step 58j if necessary.
f. From the OMNIC collect drop-down menu select Experiment Setup,
select the Bench tab and click OPEN.
Select
iN5_TransmissionMCT_Default.EXP or iN5_TransmissionTGS_Default.EXP
depending on the detector installed and click OPEN.
g. Focus the condenser for best IR throughput then close Experiment
Setup.
h. From the OMNIC View drop-down menu select Configure System Status
and then select ValPro Qualification.
i. Click Run and select the appropriate tests for the iN5 configuration.
Click OK to the settings message and then restart ValPro.
MCTA and Ge Tip ATR with MCTA
DTGS