Installing the Nicolet iN5 Microscope
This document contains confidential or proprietary informat ion of Thermo Fisher Sc ientific. Ne ither this document nor
the information therein is to be reproduced, distributed, used or disclosed, either in whole o r in part, e xcept as
specifically authorized by Thermo Fisher Scientific.
1/4/2018
Page 46 of 84
When viewing the sample aperture image in Microview for both camera only
or trinoc/camera viewer options, the sample aperture image should be
centered on the red crosshair, if not, the aperture center will be defined using
Video Calibration later in this procedure.
Sample aperture centered Sample aperture not centered
28. Open OMNIC and load the experiment file.
a. From the desktop, open OMNIC by clicking the OMNIC icon and
acknowledge main bench messages.
b. From the Collect drop down menu select Experiment Setup and click
OPEN at the bottom-left of the window.
c. Select iN5_TransmissionMCT_Default.EXP or
iN5_TransmissionTGS_Default.EXP depending on the detector installed
and click OPEN.
Refer to the following page.