RSL 5000digital

RSL 4500digital

SLIT LAMP

INSTRUCTIONS FOR USE

Digital IFU Supplement

Summary of Contents for RSL 4500digital

Page 1: ...RSL 5000digital RSL 4500digital SLIT LAMP INSTRUCTIONS FOR USE Digital IFU Supplement ...

Page 2: ...OCK RSL 5000 SLIT LAMP 9 4 2 RODENSTOCK RSL 4500 SLIT LAMP 9 5 CONTROLS 10 5 1 BACKGROUND ILLUMINATION ADJUSTMENT 10 5 2 EXPOSURE AND REVIEW BUTTONS 10 5 3 CAPTURE BUTTON 10 5 4 KEYBOARD SHORTCUTS 10 6 WARRANTY 11 7 SPECIFICATIONS AND ELECTRICAL RATINGS 11 7 1 ELECTROMAGNETIC EMISSIONS 11 7 2 ELECTROMAGNETIC IMMUNITY 12 7 3 RECOMMENDED SAFE DISTANCES 13 8 TECHNICAL SPECIFICATIONS 14 9 DIGITAL ACCE...

Page 3: ... examination of the anterior eye segment from the cornea epithelium to the posterior capsule It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment Brief description of the instrument This Rodenstock Slit Lamp can either be mounted onto a custom table top supplied by Rodenstock or can be mounted on a third parties table top refr...

Page 4: ... is used only in the configuration as supplied Should the instrument suffer shocks for example should it accidentally fall and the optical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair Do not touch accessible connectors and the patient simultaneously The owner of the instrument is responsible for training personnel in its ...

Page 5: ...DISINFECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do not autoclave or immerse in cleaning fluids Always disconnect power supply from source before cleaning 1 Wipe the external surface with a clean absorbent non shedding cloth dampened with de ionis...

Page 6: ... ADJUSTMENT long end short end RSL 5000 SLIT LAMP RSL 4500 SLIT LAMP 6 Connect the shorter USB cable 3020 P 7028 to the socket on the underside of the DCA and to the appropriate socket on the USB hub located on the Slit Lamp base Ensure that there is plenty of slack at the USB hub end to allow Slit Arm rotation 7 Route the cable in the groove in the Slit Lamp arm and replace the cover plate 8 Conn...

Page 7: ...is instrument Any servicing or repairs should only be carried out by Rodenstock Instruments or by suitably trained and authorised distributors Service manuals will be available to authorised Rodenstock service centres and Rodenstock trained service personnel 7 SPECIFICATIONS AND ELECTRICAL RATINGS The Rodenstock Digital Slit Lamp is a medical electrical instrument The instrument requires special c...

Page 8: ...t of the Digital Slit Lamp including cables than the recommended separation distances calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF IEC 61000 4 6 6 Vrms 150kHz to 80MHz 6 V d 1 2 p Radiated RF IEC 61000 4 3 10 V m 80MHz to 2 7GHz 10 V m d 1 2 p 80MHz to 800 MHz d 2 3 p 800MHz to 2 7GHz Where p is the maximum output power ra...

Page 9: ... system Microsoft Windows 10 Pro Windows 10 Pro Recommended screen resolution 1920 x 1200 pixels Software requirements CE compliant or FDA cleared camera imaging software Weight packed approx Digital Slit Lamp complete 20 0Kg 75 x 54 x 45cm W x D x H Protection against ingress IPxO Class II ME equipment Insulation between mains parts and the functional earth provide at least two means of protectio...

Page 10: ...Handle EP79 80070 1 27mm A F ALLEN KEY EP39 80329 1m DSL to PSU Cable EP39 80321 5mm L Shaped Hexagonal Key 3020 P 7077 6mm L Shaped Hexagonal Key 3020 P 5221 Black Patient Handle 3020 P 5039 Cable Clip 3020 P 7054 Chin Rest Pins 406 EP39 80365 Digital Base to Computer Cable EP39 80364 DSL Accessory Kit Packaging EP39 80250 DSL Digital Background Illum Assy 3020 P 7014 DSL Digital Backlight Filter...

Page 11: ...he environmental impact of WEEE Waste Electrical Electronic Equipment and minimise the volume of WEEE entering landfills we encourage at product end of life that this equipment is recycled and reused If you need more information on the collection reuse and recycling then please contact B2B Compliance on 01691 676124 44 1691 676124 UK only Any serious incident that has occurred in relation to the d...

Page 12: ...800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 39800 Issue A Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Straße 21 90427 Nürnberg Germany Tel 49 9131 777 1 777 Fax 49 9131 777 1 20 Visiometrics S L Vinyals 131 08221 Terrassa Spain ...